A pion beam was obtained by inserting a lead plate to absorb the photons followed by magnetic optics to select particles of equal energy. As will be shown below the beam had an electron contamination of below 1.4%.
An electron beam was obtained by inserting a copper plate into the beamline where e +e - pairs are formed in photon conversions. The magnetic optics following the converter again selected particles of equal energy. The optics was optimised for particles with production point at the copper plate, but still the electron beam had a pion contamination of nearly 50%.
For both the electron and the pion beam external particle identification
was required to obtain cleaner electron and pion samples. Three
devices were present for this in the testbeam: far upstream a
Cherenkov detector suitable for electron/pion separation up to
energies of 50 GeV; behind the TRT prototype a presampler; and just
behind that a lead glass calorimeter. The presampler was simply a
scintillator with a lead plate placed just in front of it. The signals
in the Cherenkov counter, the lead glass calorimeter and the preshower
are shown in fig. 5.9 for 20 GeV electrons and pions.
![]() |
In fig. 5.10 is shown a scatter plot of the signal in the
Cherenkov counter and the lead glass calorimeter for a mixed beam of
20 GeV electrons and pions. The regions used for the selection of
electron and pion samples are indicated.
![]() |
The efficiency of finding electrons using the particle identification
cuts for pions
and the efficiency for finding
pions using the particle identification cuts for electrons
are the two important numbers to determine.
is best measured in a beam of dominantly
electrons and an upper limit can be calculated assuming the signals in
the Cherenkov counter and the calorimeter give independent
measurements. The efficiency for finding electrons when making cuts
for pions is
|
| (77) |
Four different selection criteria can be defined:
|
Xtotal = Xe + X | (78) |
Since the cuts on the Cherenkov counter and the calorimeter are independent
|
| (79) |
|
| (80) |
The analysis of
and
<
(
)
(
) are similar
to the calculation of
and the results are
An upper limit on the amount of electrons in the pion beam before the
particle identification, is given by the fraction of events passing
selection criteria D in the pion beam scaled with the fraction of
electrons actually passing this selection criteria (estimated from the
electron beam). The results are that the pion beam at the most
contains 1.4% electrons and the electron beam at the most contains
60% pions.
After the external cuts both beams will be pure to below the level of
10- 4 which is sufficient for the electron identification studies
where rejections below
5
10- 3 are never reached. The results
are summarised in table 5.2
| Maximum fraction | Maximum fraction | |
| before selection | after selection | |
| electrons in pion beam | 0.014 |
7.3 |
| pions in electron beam | 0.603 |
2.1 |