Run with pedestal subtraction and zerosuppression ------------------------------------------------- Pedestal and rms for each channel: chamber-fec01-swtrig-clk0-ped0-thr0-g0-s0-p1-1008-4-pedrun.csv.pdf Normalized timesweeps for pedestal run: norm.pdf (voltage on chamber) norm-novolt.pdf (voltage off chamber) blue: average of all events green: max for each sample of all events red: min for each sample of all events Timesweeps for all channels for an individual event: timesweeps-ped-evt[nb>1].pdf Pedestal subtraction and zerosuppression(thrreshold=5): allevents (20): chamber-fec01-zsup-clk0-ped1-thr5-g0-s0-p1-1008-3-evt0.pdf individial events: chamber-fec01-zsup-clk0-ped1-thr5-g0-s0-p1-1008-3-evt[nb>0].pdf Timesweeps for all channels for an individual event: timesweeps-evt[nb>1].pdf Conclusion ---------- MCM0: there is a disturbance (, not for all events and (perhaps not/perhaps) synchronized with trigger. MCM1: on average a "small" wavy structure.