Data without zero suppression: all-twochannels.pdf =================== Data taken with SRU and protoMCM. 915 samples and two channels at time. Pulser with cirka 20kHz, which induce a pulse on the input on the protoMCM. The plots show histogram of all the ADC values around the peak for each channel. Cirka 150000 events. all-ch0-15.pdf ============== Data taken with SRU and protoMCM. 115 samples and all channels. Pulser with cirka 20, 40 or 80 kHz (started with 20 -> 80), which induce a pulse on the input on the protoMCM. The plots show histogram of all the ADC values for each channel. Cirka 56000 events. Not all events have a pulse due the short sampling time. One can see some structure in that some ADC values are preferred. Due to SALTRO,MCM,pulser,clock? No further analysis done yet. Took also data with zerosuppression: threshold 30 and 915 samples, but data is sometimes corrupt. Could see that some channels do not have any signal. all-ch0-15-zsup.pdf =================== Used clusters with maxADC > 200. Some channels have several lowADC clusters, do not know why yet. For some runs was it possible to only take exactly 28160 (in hex 6E00) events, then the SRU/CPLD hanged and had to be reset. It also had intermittent hangups, but then it mostly timed out and the run could continue.